Early Hemodynamic findings of St Jude Medical Trifecta Pericardial Valve
Hyung Tae Sim, Hyoung Woo Chang, Jae Suk Yoo, Kwang Ree Cho, Dong Jin Kim
Department of Thoracic and Cardiovascular Surgery, Sejong General Hospital, Gyeonggi-do, Korea

Background : The study objective was to assess the results of early hemodynamic findings of St Jude Medical Trifecta valve

Methods : A review was conducted of 75 patients (M:F = 32 : 43) who underwent aortic valve replacement with SJ Trifecta medical valve between December 2014 to July 2016. Indication for aortic valve replacement surgery included pure stenosis in 32 patients, regurgitation in 8 patients, mixed pathology in 27 patients and others in 8 patients. The number of used valve sizes are as follows; 19mm: 32, 21mm: 24, 23mm: 9, 25mm: 5, 27mm:5.

Results : Early in hospital mortality was 5.3%(4/75). At the time of discharge average ejection fraction was 52.5 ± 11.9, mean aortic pressure gradient was 11.2 ± 5.5 mmHg and average peak aortic pressure gradient was 20.3 ± 10.2 mmHg respectively. There was no paravalvular leak observed.

Conclusion : The St Jude Medical Trifecta valve is a unique pericardial bioprosthesis with design elements that provide tolerable hemodynamic performance in early post operative period